Medical X-ray Diagnostic Radiation Protection Requirements

Medical X-ray Diagnostic Radiation Protection Requirements

Medical X-ray Diagnostic Radiation Protection Requirements (GBZ 130-2013)
#1
According to the work content, each X-ray equipment should be equipped with not less than the basic types of staff, subject protective equipment and auxiliary protective facilities required in Table 4. The number should meet the needs of the work, and at least lead Rubber protective clothing.

#2
■ The requirements for the configuration of personal protective equipment and auxiliary protective facilities in the equipment room of on-board diagnostic X-ray equipment shall be implemented according to the type of equipment installed in Table 4.

#3
■ Except for interventional protective gloves, the lead equivalent of protective equipment and auxiliary protective facilities should not be less than 0.25 mmPb; the lead equivalent of interventional protective gloves should not be less than 0.025 mmPb; the lead equivalent of thyroid and gonad protective equipment should not be less than 0.5 mmPb; mobile lead protective screens Lead equivalent should not be less than 2mmPb.

#4
■ The X-ray examination of children should be equipped with protective equipment to protect the corresponding tissues and organs, and the lead equivalent of the protective equipment and auxiliary protective facilities should not be less than 0.5mmPb.

#5
■ When personal protective equipment is not in use, it should be stored properly and should not be folded to prevent breakage.

#6
■ For places where mobile X-ray equipment is frequently used (such as: intensive care, critically ill patient treatment, orthopaedic reduction, etc.), a sufficient number of mobile lead protective screens should be provided.